auto chemistry analyzer
Model:EM120
SKU00035
-
EM120 Auto Chemistry Analyzer: Key Features
- High Throughput: Achieves 120 tests/hour with dual-reagent capability for efficient workflow.
- Ample Reagent Capacity: 80 refrigerated positions minimize reloading and streamline operations.
- Smart Detection: Automatic liquid-level sensing and anti-collision protection ensure precision and probe safety.
- Advanced Optics: Integrated full-closed optical system eliminates contamination for reliable, accurate results.
- Premium Reaction Cups: New amorphous optical plastic cuvettes ensure high clarity and consistent readings.
- Low Maintenance: Semiconductor solid-state refrigeration (Peltier cooling) requires no upkeep.
- Intuitive Design: User-friendly interface with ergonomic controls for seamless operation.
- Hassle-Free Workflow: Disposable cuvettes eliminate cleaning and reduce cross-contamination risks.
- Quality Assurance: Rigorous ISO-certified standards guarantee dependable performance.
- Cost-Effective: Balances productivity and accuracy with minimal operational costs.
Ideal for labs prioritizing speed, precision, and affordability!
EM120 Auto Chemistry Analyzer: Technical Specifications
General Specifications
- Machine Type: Random access, open reagent system (locked reagent optional).
- Test Speed: 120 tests/hour (mono or dual reagent).
- Test Principles: Colorimetric method, turbidimetry.
- Test Methods:
- 1-point end, 2-point end.
- Fixed-Time, Kinetics.
Sample System
- Sample Tray:
- 40 positions (supports micro cups, test tubes, blood collection tubes).
- Sample Probe:
- Liquid surface detection.
- Collision avoidance.
- Sample Volume: 2–30μL (adjustable in 0.1μL increments).
- Optional: Barcode reader for sample identification.
Reagent System
- Reagent Tray:
- 80 positions (40 for R1, 40 for R2).
- Non-stop Peltier cooling (2°C–14°C).
- Open system (supports third-party reagents).
- Reagent Volume: 20–300μL (adjustable in 1μL increments).
- Reagent Probe:
- Liquid surface detection.
- Collision avoidance.
Reaction System
- Reaction Cuvettes: 50 cuvettes (amorphous optical plastic for high transmittance).
- Reaction Volume: 150–330μL.
- Reaction Time: 10 minutes.
- Temperature Control: Peltier incubation system (37°C ±0.1°C).
Optical System
- Light Source: Halogen lamp.
- Wavelengths:
- Standard: 340nm, 405nm, 450nm, 505nm, 546nm, 578nm, 630nm, 700nm.
- Optional: 4 additional wavelengths (customizable).
- Absorbance Range: 0–4.0 Abs.
- Resolution: 0.0001 Abs.
Mixing System
- Mixer Type: Paddle-style mixer.
- Features: Reduces carry-over for contamination-free results.
Calibration & Quality Control
- Calibration Types:
- Linear and nonlinear.
- Multi-point calibration (up to 8 points).
- Quality Control:
- Supports Westgard rules, Levey-Jennings (L-J) plots, cumulative QC.
- Automatic result checking and error alerts.
Software & Data Management
- Operating System: Compatible with Windows XP/7/8/10.
- Features:
- Automatic calibration and dilution (decrease/increase/normal modes).
- Serum index, blank deduction, cross-contamination prevention.
- Patient data management, report auditing, fuzzy query, and statistics.
- Customizable report formats (6 templates + user-defined options).
- User privilege management and alarm logs.
Interfaces
- Network: TCP/IP connectivity.
- Optional: Reagent barcode reader.
Physical Specifications
- Dimensions: 750mm × 550mm × 490mm (L×W×H).
- Weight: 70kg.
- Power Supply: AC 110/220V ±10%, 50/60 Hz, 650W.
- Water Consumption: 6L/hour (max).
Key Advantages
- Flexibility: Open reagent system with Peltier-cooled storage.
- Precision: High-resolution optical system and temperature control.
- Efficiency: Automated workflows with advanced QC and calibration.