chemistry analyzer
Model:EM380
SKU00052
EM380 Auto Chemistry Analyzer: Technical Specifications
Key Features
- Smart Design: Innovation-focused engineering for high performance and reliability.
- Test Speed: 300 tests/hour ensures efficient workflow and time-saving operations.
- Automated Testing & Low Maintenance: Streamlined processes with minimal manual intervention, reducing downtime.
- Reagent Consumption: 150μL per test optimizes cost-efficiency without compromising accuracy.
EM380 Auto Chemistry Analyzer: Comprehensive Specifications
Key Advantages
-
Top Precision:
- Advanced sampling technologies with automatic probe depth adjustment.
- Sensitive liquid sensors for real-time monitoring of reagent/sample residuals.
- High-precision ceramic syringe for accurate dispensing (as low as 0.1 µL).
-
Best Quality:
- Non-stop Peltier cooling system (2–14°C) for reagent stability.
- Teflon-coated mixer reduces carry-over contamination.
- Triple-wash station ensures clean results and valid diagnostics.
-
Cost-Effective:
- Low reagent consumption (150 µL per test).
- Durable ceramic piston and reusable reaction cuvettes minimize operational costs.
-
Service:
- Comprehensive support for seamless operation and maintenance.
- Streamlined workflows enhance laboratory productivity.
Technical Specifications
1. Sampling Probe
- Design:
- External and internal mirror polish for reduced carry-over.
- Dedicated internal/external probe washing.
- Features:
- Liquid sensor for real-time residual feedback.
- Collision protection and automatic depth adjustment.
2. Multi-Functional Sample & Reagent Tray
- Sample Tray:
- 70 positions (supports micro cups and test tubes).
- Reagent Tray:
- 70 refrigerated positions (2–14°C, Peltier cooling).
- Optional:
- Barcode reader for sample/reagent tracking.
3. Reaction System
- Reaction Tray:
- 90 disposable cuvettes.
- Reaction volume: 150 µL (minimum).
- Temperature Control:
- Maintains 37°C ± 0.1°C for optimal reaction stability.
4. Mixer Design
- Mixing Technology:
- Teflon-coated paddle mixer.
- Prevents water droplets and ensures uniform mixing.
5. Washing Station
- Design:
- 5-needle wash station.
- Triple washing process for thorough cleaning.
6. High-Precision Ceramic Syringe
- Dispensing:
- Frequency-converting design for accuracy.
- Permanent ceramic piston for durability.
- Dispensing range: 0.1–300 µL.
Additional Features
- Optical System:
- Halogen lamp with multiple wavelengths for photometric accuracy.
- Software:
- User-friendly interface with real-time monitoring and error alerts.
- Dimensions:
- Compact design for efficient lab space utilization.