Manual WBC Chamber Cleaning & HGB Alarm Resolution: Step-by-Step Video guide
2025-03-05 16:17
HGB报警解决办法
1. 先点击服务进入服务办面后,再点自检进入自检界面。
Click "Service" to enter the service office, and then click "self-check" to enter the self-check interface.
2. 进入自检界面后点击HGB调整(HGB Adj)后显示4.2正负0.5为正常 。
After entering the self-check interface, click HGB Adjustment (HGB Adj) and 4.2 + / - 0.5 is displayed
3. 如果第2步显示不为4.2正负0.5,请双击以下视频观看(Manually Clean WBC Chamber(English)_210708_01),按照视频中的操作步骤清洗完WBC样品杯后,再执行第1和第2步。
If step 2 is not 4.2 plus or minus 0.5, double-click the following video to watch (Manually Clean WBC Chamber(English)_210708_01) and then perform steps 1 and 2 after cleaning the WBC sample cup according to the steps in the video