EM6100 Hematology Analyzer SOP: Step-by-Step Operation & Maintenance Guide
Standard Operating Procedure (SOP) – Analyzer Maintenance & Operation
1. Power Requirements
-
A. Grounding:
-
Ensure proper earth grounding before use.
-
-
B. Fuse & Power Cord:
-
Use only the specified fuse type and provided power cord.
-
-
C. Voltage Check:
-
Verify input voltage matches analyzer requirements before powering on.
-
2. Environmental Requirements
-
A. Temperature:
-
Normal range: 15°C – 30°C
-
Operating range: 10°C – 40°C
-
-
B. Humidity:
-
Normal range: 20% – 85% RH
-
-
C. Atmospheric Pressure:
-
70 kPa – 106 kPa
-
-
D. Workplace Conditions:
-
Free from dust, vibration, contamination, noise, and electrical interference.
-
Avoid strong electromagnetic fields (e.g., brush motors, flickering lights, frequent electrical switching).
-
No direct sunlight, heat sources, or drafts.
-
Well-ventilated area.
-
-
E. Placement:
-
Use on a stable, level indoor workbench.
-
Connect only to a properly grounded outlet.
-
3. Pre-Operation Checks
-
A. Reagent Inspection:
-
Check for sufficient volume, turbidity, deterioration, and bent tubing.
-
-
B. Connections:
-
Verify correct external device and power connections.
-
4. Startup & Login
-
Turn on the printer (if connected).
-
Power on the analyzer.
-
Enter username and password at the login prompt.
5. Background Check (Automatic on Startup)
-
Acceptable Limits:
-
WBC ≤ 0.2×10⁹/L
-
RBC ≤ 0.02×10¹²/L
-
HGB ≤ 1g/L
-
PLT ≤ 5×10⁹/L
-
HCT ≤ 0.5%
-
-
If Failed:
-
A second retest occurs automatically.
-
If still failed, an "Abnormal Background" alarm triggers—refer to the Troubleshooting manual.
-
6. Daily Quality Control (QC)
-
Run QC tests before sample analysis (refer to Chapter 7: QC in the manual).
7. Selecting Work Mode
-
Options:
-
Whole Blood Mode
-
Pre-Diluted Mode
-
-
Sample Input Methods:
-
Enter all information manually or
-
Input only the Sample ID.
-
-
Click "OK" to confirm settings.
8. Sample Analysis
Whole Blood Mode
-
Mix EDTA-anticoagulated sample thoroughly.
-
Place under the probe and press [Aspirate].
-
Remove tube after beep; analysis begins automatically.
Pre-Diluted Mode
-
Click "Dilute" and follow on-screen instructions.
-
Press [Aspirate] to add 200μL diluent (avoid bubbles).
-
Add 20μL blood, cap, and mix.
-
Wait 5 minutes, remix, then aspirate for analysis.
9. Standby Mode
-
Activates after idle time (set in settings).
-
Press [Aspirate] to resume operation.
10. Shutdown Procedure
-
Click "Shutdown" on the main menu.
-
The analyzer performs probe cleaning.
-
At "Confirm" prompt, click "OK" and power off all devices.
-
Empty and properly dispose of waste containers.
11. Troubleshooting
-
Click the error message (bottom-right of screen) for guidance.
-
Refer to Chapter 11: Troubleshooting in the manual.
12. Regular Maintenance
Daily:
-
Normal shutdown with probe cleaning.
Monthly:
-
Wipe external surfaces and probe wipe block (use 75% alcohol).
Extended Non-Use (>2 Weeks):
-
Perform Pack-up function and power off the analyzer.