EM Series Chemistry Analyzer: Probe Collision Prevention & Adjustment Guide
2025-04-18 11:23
How to Prevent Probe Collision in the EM Series Fully Auto Chemistry Analyzer
To ensure smooth operation and avoid probe damage, follow these guidelines:
1. Proper Cuvette Usage
- Do not use the top cuvette position when two cuvettes are placed incorrectly
- Incorrect placement may cause the needle to hit the cuvette bottom, leading to collision and probe damage
2. Avoid Unauthorized Adjustments
- Do not adjust settings without instruction — this can disrupt factory calibration
- Adjustments should only be made by our certified engineers to maintain precision
3. Critical Adjustment Positions
Adjustment Direction for All Positions:
• Clockwise rotation → Increases value
• Counterclockwise rotation → Decreases value
• For optimal performance, set to the middle position
• Clockwise rotation → Increases value
• Counterclockwise rotation → Decreases value
• For optimal performance, set to the middle position
1. Washing Position Adjustment

Washing Position Mechanism
Adjust to ensure proper probe cleaning without contact
2. Reagent Position Adjustment


Ensures accurate reagent aspiration without collision
3. Cuvette Position Adjustment

Cuvette Position Mechanism
Critical for proper sample measurement alignment
Important Adjustment Notes:
- The same adjustment logic applies to all three positions
- Always verify alignment after making adjustments
- For first-time adjustments, contact our technical support