How to Prevent Probe Collision in the EM Series Fully Auto Chemistry Analyzer
To ensure smooth operation and avoid probe damage, follow these guidelines:
1. Proper Cuvette Usage
- Do not use the top cuvette position when two cuvettes are placed incorrectly
- Incorrect placement may cause the needle to hit the cuvette bottom, leading to collision and probe damage
2. Avoid Unauthorized Adjustments
- Do not adjust settings without instruction — this can disrupt factory calibration
- Adjustments should only be made by our certified engineers to maintain precision
3. Critical Adjustment Positions
• Clockwise rotation → Increases value
• Counterclockwise rotation → Decreases value
• For optimal performance, set to the middle position
1. Washing Position Adjustment

2. Reagent Position Adjustment


3. Cuvette Position Adjustment

- The same adjustment logic applies to all three positions
- Always verify alignment after making adjustments
- For first-time adjustments, contact our technical support