Phone +86-755-33275355 Email info @ east-medical. com
Home / News / EM Series Chemistry Analyzer: Probe Collision Prevention & Adjustment Guide

EM Series Chemistry Analyzer: Probe Collision Prevention & Adjustment Guide

2025-04-18 11:23

 

How to Prevent Probe Collision in the EM Series Fully Auto Chemistry Analyzer

To ensure smooth operation and avoid probe damage, follow these guidelines:

1. Proper Cuvette Usage

  • Do not use the top cuvette position when two cuvettes are placed incorrectly
  • Incorrect placement may cause the needle to hit the cuvette bottom, leading to collision and probe damage

2. Avoid Unauthorized Adjustments

  • Do not adjust settings without instruction — this can disrupt factory calibration
  • Adjustments should only be made by our certified engineers to maintain precision

3. Critical Adjustment Positions

Adjustment Direction for All Positions:
Clockwise rotation → Increases value
Counterclockwise rotation → Decreases value
• For optimal performance, set to the middle position

1. Washing Position Adjustment

Washing Position Adjustment
Washing Position Mechanism
Adjust to ensure proper probe cleaning without contact

2. Reagent Position Adjustment

 

Reagent Position Adjustment
Ensures accurate reagent aspiration without collision

3. Cuvette Position Adjustment


Cuvette Position Mechanism
Critical for proper sample measurement alignment
Important Adjustment Notes:
  • The same adjustment logic applies to all three positions
  • Always verify alignment after making adjustments
  • For first-time adjustments, contact our technical support

Why Choose Our Auto Chemistry Analyzer?

  • Reliable Performance – Precision-engineered for accuracy
  • 2-Year Warranty – Full coverage for peace of mind
  • Expert Support – Dedicated technical assistance
Need professional assistance with adjustments? Contact our support team today.