Introduction to Semi-Auto Biochemistry Analyzers
In the world of modern medical diagnostics, Semi-Auto Biochemistry Analyzers step into the spotlight as indispensable tools. Often referred to as Semi Auto Analyzers, these
Redefining in-vitro Diagnostics we strive to improve care and outcomes
through inspired and world-leading healthcare solutions
Home » NEWS & EVENTS
In the world of modern medical diagnostics, Semi-Auto Biochemistry Analyzers step into the spotlight as indispensable tools. Often referred to as Semi Auto Analyzers, these
Daily maintenance After every day’s worktime. 1, Function-Service-Maintenance-Maintain-Flush 2, Burn WBC Aperture & RBC Aperture 3, Inverted Flush WBC Aperture & RBC Aperture Weekly maintenance
Power on 1.1 After power on, the machine will confirm whether the blank count is qualified: WBC≤0.3, RBC≤0.03, HGB≤1, PLT≤10. (Each time the instrument is
EM6100 three-part hematology analyzer Standard Operating Procedure A. Power requirement B. Environment requirement Operation Guide and maintenance 1. Checking before powering on Check whether the reagent is sufficient, turbidity and deterioration, and the reagent tube is bent. Check whether the analyzer is connected to external device correctly.Check whether the power connection is correct. 2. Startup and Login If the analyzer connects to printer, turn on the printer switch. Then turn on the power switch at the back
Power-On Procedure 1. Reagent Pack Installation: 2. Pump Tube Placement: 3. Electrode Installation: 4. Printer Paper: 5. Power-On: 6. Self-Check: Sample Preparation 1. Hemolysis Check:
Upgrade your lab’s productivity with Eastmed’s double chamber hematology analyzer. This cost-effective solution offers a 60 samples per hour throughput, a low maintenance design, and a user-friendly 10.4 inch touch screen for reliable performance.
Hematology analyzers are instruments used to analyze blood samples to determine the number and types of cells present in the blood. A 3-part hematology
Offset value adjust Adjust the optical offset value(the signal when there is no light through the flowcell) The steps about adjusting the optical offset value:
Hematocrit (HCT): normal range males 40% to 54%; females 37% to 47%; neonates 48% to 68%. Mean red blood cell volume (MCV): refers to the
what is the kinetic reaction method? The kinetic reaction method is an analysis method used in clinical chemistry to determine the difference in absorbance between
EM 200plus is a fully automated, random access and clinical chemistry analyzer that enhances productivity and turnaround time. It has a throughput of 200 photometric